ULTRASONİK BANYO - SONİKATÖR

ULTRASONİK BANYO - SONİKATÖR

 

ULTRASONİK BANYO

 

SUPER RK …

 

DIGITEC DT …

DIGIPLUS DL …

Capacity (L)

0.9 – 90.0

0.9 – 90.0

3.0 – 28.0

Time setting (min)

1–15, ± 5 %

continuous operation (∞)

1, 2, 3, 4, 5, 10, 15, 30, ± 5 %

continuous operation (∞)

1, 2, 3, 4, 5, 10, 15, 30, ± 5 %

continuous operation (∞)

Safety shut-down

No

after 12 hours

After 12 hours

Heating

Optional, version „H“

Optional, version „H“

3

Adjustable temp.range (oC)

30 – 80

RK 31 H : 65 fixed

20 – 80

20 – 80

 

Excess temp. signal

No

Yes

Yes

Protection aganist boiling retardation

No

Yes

Yes

Setting accuracy of bath temp. (K)

± 5 K

± 2.5 K

± 2.5 K

Thickness tank (mm) / material C-version :

0.8 / 1.4301

2.0 / 1.4404

0.8 / 1.4301

2.0 / 1.4404

0.8 / 1.4301

 

Filling mark for safe dosage

Yes

Yes

Yes

Hard chromium-plated

RK 102 H

DT 102 H / H-RC

DL 102 H

One-piece drain, welded

Yes , from RK 102 H

Yes , from DT 102 H

Yes

Degree of protection

IP 32

IP 33

IP 33

Ultrasonic frequency (kHz)

35

35

35

SweepTec

Yes

Yes

Yes

Pulse function

Yes

Yes

Yes

Power setting

No

No

20 – 100 % in 10 % steps

Oscillating systems

Yes

Yes

Yes

Fast degassing DEGAS

No

Yes

Yes

Mains supply:

230 V - (+- 10%) 50/60 Hz

115 V – (+-10%) 50/60 Hz

Yes

Yes

Yes

Data memory

No

No, Type H-RC: WINSONIC®software

No

 

Interface / PC software

No

RS 232 bei Typ H-RC /3

No

Medical device class I

Yes

Yes

No

 

ULTRASONİC SONİKATÖR

 

SERİE 4000

 

MİNİ20

SERİE 2000.2

 

ADVANCED

SPECİAL

BASIC

Sample volumes Batch Flow-through

0.5 to 1000 ml.

Up to 30 l / h

0.1 to 25 ml.

1 to 1000 ml.

Up to 30 l / h

Ultrasonic converter

Possible configurations:

GM 4100 with UW 50 or

UW 100 and SH 100 G

 

GM 4200 with UW 50 or

UW 200 with SH 200 G

 

GM 4400 with UW 400 and SH 400 G or with UW 200 and SH 200 G

1

1

Amplitude setting

10 to 100 %

10 to 100 %

10 to 100 %

Automatic amplitude limiting

After input of the assembled probe

After input of the assembled probe

After input of the assembled probe

Pulsation

ON cycles 0.2 – 600 s

OFF cycles 0.3 – 600 s

ON cycles 0.1 – 60 s

OFF cycles 0.2 – 60 s

ON cycles 1 – 60 s

OFF cycles 1 – 60 s

Time setting

9 h : 59 min : 59 s

Continuous or timed

 

59 min : 59 s

59 min : 59 s

Continuous or timed

 

Safety shut-down

9 h : 59 min : 59 s

59 min : 59 s

9 h : 59 min : 59 s

Indicators

Alphanumerical LC display of amplitüde, pulse level, time, energy input, temperature (optional)

Alphanumerical LC display of amplitüde, pulse level, time, energy input

Alphanumerical LC display of amplitüde, pulse level, time

Energy display

in kJ

in kJ

No

Temperature display and measurement

Optional, 0 to 120 oC,

Temperature sensor required, alternatively beep or switch off

No

No

Batch operation Sequencing

Yes,

Several batches one after the other

No

No

Remote control with PC

RS 232 (Sub-D)

RS 232 (infrared)

No

Fault diagnosis

Yes

Yes

Yes

Operating frequency

20 kHz

30 kHz

20 kHz

Program memory

Yes, 9

Yes , 9

No

Functional test

Yes

Yes

No

Mains connection

230 V-(+- 10%),

Alternatively 115 V-(+- 10 %), 50/60 Hz

100-240 V- , 50 / 60 Hz

230 V-(+- 10%),

Alternatively 115 V-(+- 10 %), 50/60 Hz